MCD Series
| model : MCD-05-05002 | Multi-channel Charge/Discharge Tester |
![]() |
MCD series is designed to emulate various batteries such as secondary batteries, EDLC, Coin cells as well as HILS and ECU battery monitoring modules.
Up to 50 channels of synchronized operation is possible while all channels are isolated. All channels are independently controlled accurately.
With the hardware sequence, it is enabled to set as fast as 2μs max and 3ms measurement time. Optional low cost software is available.
- 5 channels of charger/discharger in 1U chassis
- Expandable to max 50 channels (Hardware synchronization)
- Expandable to max 200 channels (Software synchronization)
- Various measurement (Terminal voltage, charge/discharge current, capacitance and so on.)
- Ease of operation by dedicated software
MCD-05-05002 can be connected in paralleled or in series as to meet various applications such as secondary battery evaluation or stacked battery output simulation.
MCD-05-05002 is incorporated 5 channels in 1 unit. Each channel is isolated.
By the master & slave configuration, up to 50 channels can be configured for multichannel applications.
| MCD-05-05002 | 5 ch. of 5V, 200mA |
| MCD-05-05002/REC | Test Data for MCD-05-05002 |
| Control PC | ||
|---|---|---|
| No. of Channel | 5CH/1, max50CH/10(hardware connection), max200CH/40(software connection) | |
| Function | Charge | Charge (DC V & A output) |
| Discharge | Discharge (DC A, Loading ) | |
| Measurement | DC V & A and by calculation | |
| Setting | I/F | Through control PC using USB I/F (No setting on the panel) |
| Mode | Single CH/All CH setting (each CH setting error max 10ms) | |
| Operation | Real time operation | Control by real time commands from control PC |
| Sequence operation | Sequence to be set in the control PC | |
| Control PC | Hardware | IBM PC-AT or equivalent |
| Performance | CPU: Pentium 4.2GHz or better | |
| Memory: 2GByte or more | ||
| HDD capacity for program: 1GByte or more | ||
| OS | Microsoft WindowsXP, Windows Vista | |
| Charge | ||
| Function | Charge (DC voltage/current output) | |
| Operation mode | CC mode, CV mode (automatic switching) | |
| Protector | Over current | By firmware using setting and reading |
| Over voltage | By firmware using setting and reading | |
| Output voltage | Setting range | 5V to 0V (single range) |
| Setting accuracy | ±0.06% of f.s. *1 | |
| Setting resolution | 1mV | |
| Output current | Setting range | 200mA to 0mA (single range) |
| Setting accuracy | ±0.075% of f.s. | |
| Setting resolution | 0.1mA | |
| Discharge | ||
| Function | Discharge (DC A, Loading) | |
| Operation mode | CC mode (single mode) | |
| Protector | Over current and voltage | By firmware using setting and reading |
| Output voltage | Setting range | 5V to 0V (single range) |
| Setting accuracy | ±0.06% of f.s. *1 | |
| Setting resolution | 1mV | |
| Load current | Setting range | 0 to -200mA (single range) |
| Setting accuracy | ±0.075% of f.s. | |
| Setting resolution | 0.1mA | |
| Measurement | ||
| Measurement mode | Terminal voltage of DUT | DC V |
| Charging current | DC A | |
| Discharge current | DC A | |
| Static capacitance of capacitor | By software | |
| Battery capacitor | By software | |
| Internal Resistance | By software | |
| DC V | Range | 6V to 0V (single range) |
| Accuracy | ±0.02% of f.s. *2 *3 | |
| Resolution | 0.1mV | |
| DC A | Range | 220mA to -220mA (single range) |
| Accuracy | ±0.05% of f.s. *2 *4 | |
| Resolution | 0.1mA | |
| Time | Resolution | ±0.2% of rdg. |
| Accuracy | 3ms (hardware) | |
| Range | Hardware range: 3ms to 60,000ms (No upper limit when combined with software) |
|
| Interface | ||
| USB | USB1.1 | |
| I/O | 8ch photo coupler isolated OC output (12V/10mA, max24V/10mA) |
|
| Master-slave expansion | Dedicated connector | |
| General | ||
| Rated input voltage | AC100V to 240V ±10% 50/60Hz | |
| power consumption | 40VA or less | |
| Dimensions | 430(W) x 44(H) x 400(D) EIA/1U | |
| Weight | Approx. 4kg | |
| Operational temperature/humidity | 10℃ to 40℃/10% to 90%RH (no dew condensation) | |
| Accuracy guaranteed temperature & humidity range | Guaranteed for 6 months under 23℃±5℃/70% RH | |
*1 Guaranteed setting accuracy range: 0.5V to 5V
*2 Higher conversion speed gives lower accuracy. To obtain the above accuracy, keep the initial conversion speed(55Hz).
*3 Guaranteed accuracy range: 0V to 5V. The voltage FS is 5V.
*4 Guaranteed accuracy range: -200mA to 200mA. The FS is 200mA.