MCD Series
model : MCD-05-05002 Multi-channel Charge/Discharge Tester   

 

Features

MCD series is designed to emulate various batteries such as secondary batteries, EDLC, Coin cells as well as HILS and ECU battery monitoring modules.
Up to 50 channels of synchronized operation is possible while all channels are isolated. All channels are independently controlled accurately.
With the hardware sequence, it is enabled to set as fast as 2μs max and 3ms measurement time. Optional low cost software is available.

  • 5 channels of charger/discharger in 1U chassis
  • Expandable to max 50 channels (Hardware synchronization)
  • Expandable to max 200 channels (Software synchronization)
  • Various measurement (Terminal voltage, charge/discharge current, capacitance and so on.)
  • Ease of operation by dedicated software

MCD-05-05002 can be connected in paralleled or in series as to meet various applications such as secondary battery evaluation or stacked battery output simulation.
MCD-05-05002 is incorporated 5 channels in 1 unit. Each channel is isolated.
By the master & slave configuration, up to 50 channels can be configured for multichannel applications.

 

Order Information
MCD-05-05002 5 ch. of 5V, 200mA
MCD-05-05002/REC Test Data for MCD-05-05002

 

Specifications
Control PC
No. of Channel 5CH/1, max50CH/10(hardware connection), max200CH/40(software connection)
Function Charge Charge (DC V & A output)
Discharge Discharge (DC A, Loading )
Measurement DC V & A and by calculation
Setting I/F Through control PC using USB I/F (No setting on the panel)
Mode Single CH/All CH setting (each CH setting error max 10ms)
Operation Real time operation Control by real time commands from control PC
Sequence operation Sequence to be set in the control PC
Control PC Hardware IBM PC-AT or equivalent
Performance CPU: Pentium 4.2GHz or better
Memory: 2GByte or more
HDD capacity for program: 1GByte or more
OS Microsoft WindowsXP, Windows Vista
Charge
Function Charge (DC voltage/current output)
Operation mode CC mode, CV mode (automatic switching)
Protector Over current By firmware using setting and reading
Over voltage By firmware using setting and reading
Output voltage Setting range 5V to 0V (single range)
Setting accuracy ±0.06% of f.s. *1
Setting resolution 1mV
Output current Setting range 200mA to 0mA (single range)
Setting accuracy ±0.075% of f.s.
Setting resolution 0.1mA
Discharge
Function Discharge (DC A, Loading)
Operation mode CC mode (single mode)
Protector Over current and voltage By firmware using setting and reading
Output voltage Setting range 5V to 0V (single range)
Setting accuracy ±0.06% of f.s. *1
Setting resolution 1mV
Load current Setting range 0 to -200mA (single range)
Setting accuracy ±0.075% of f.s.
Setting resolution 0.1mA
Measurement
Measurement mode Terminal voltage of DUT DC V
Charging current DC A
Discharge current DC A
Static capacitance of capacitor By software
Battery capacitor By software
Internal Resistance By software
DC V Range 6V to 0V (single range)
Accuracy ±0.02% of f.s. *2 *3
Resolution 0.1mV
DC A Range 220mA to -220mA (single range)
Accuracy ±0.05% of f.s. *2 *4
Resolution 0.1mA
Time Resolution ±0.2% of rdg.
Accuracy 3ms (hardware)
Range Hardware range: 3ms to 60,000ms
(No upper limit when combined with software)
Interface
USB USB1.1
I/O 8ch photo coupler isolated OC output
(12V/10mA, max24V/10mA)
Master-slave expansion Dedicated connector
General
Rated input voltage AC100V to 240V ±10% 50/60Hz
power consumption 40VA or less
Dimensions 430(W) x 44(H) x 400(D) EIA/1U
Weight Approx. 4kg
Operational temperature/humidity 10℃ to 40℃/10% to 90%RH (no dew condensation)
Accuracy guaranteed temperature & humidity range Guaranteed for 6 months under 23℃±5℃/70% RH

*1 Guaranteed setting accuracy range: 0.5V to 5V
*2 Higher conversion speed gives lower accuracy. To obtain the above accuracy, keep the initial conversion speed(55Hz).
*3 Guaranteed accuracy range: 0V to 5V. The voltage FS is 5V.
*4 Guaranteed accuracy range: -200mA to 200mA. The FS is 200mA.